Base of Standards » Special state standard for the relative permittivity unit of liquid, solid and gaseous dielectrics in the range of 1-178,4 GHz GET 110-2012 »

    State primary standard units for the complex dielectric permittivity in the range frequencies from 1 to 78,33 GHz GET 110-2010

    Name – State primary standard units for the complex dielectric permittivity in the range frequencies from 1 to 78,33 GHz GET 110-2010

    • Purpose – Reproduction and keeping units of complex dielectric permittivity   in the range frequencies from 1 to 78,33 GHz and transferring of the units size by working standards to working means of measurements for metrological assurance in accordance with the State verification schedule of GOST R 8.711-2010
    • Certification Approval  Year – 2010
    • Rating value for solid dielectrics of:

    Relative dielectric permittivity…….1,2 – 500

    Tangent  of dielectric losses ………..10-8 – 10-2

    • Random error estimation of unit reproduction of:

    Relative dielectric permittivity……….5·10-5 – 1×10-3

    Tangent  of dielectric losses ………….1×10-2– 5×10-2

    • Residual bias estimation of unit reproduction of:

    Relative dielectric permittivity……….1×10-4– 3×10-3

    Tangent  of dielectric losses …………..1×10-2– 1×10-1

    • Standard description – GOST R 8.711-2010 «State verification schedule for measuring instruments of relative dielectric permittivity and  the loss tangent in the range of frequencies from 1 to 78,33  GHz»
    • The standard consists of:

    – four standard systems for reproduction units of the relative dielectric permittivity and tangent of dielectric losses in the frequency range of 1-78,33 GHz, starting measuring instruments of the module of factor of transfer and reflexion R2M-18 оr R2М-40 depending on a range of frequencies, frequency multipliers, a set of measuring resonators  and means of its thermostat control, reference measures of the relative dielectric permittivity and tangent of dielectric losses

    • Scope of application:

    – information telecommunication technology and electronic engineering (optoelectronics, radioelectronics, acoustoelectronics, optical and microwave communication, hardware components of microelectronics, nanoelectronics and quantum computers)

    – new materials and chemical technologies (microtechnological, nanotechnological, glass, ceramic, composition materials and polymers)

    – fundamental scientific research in condensed matter physics.

    • Scientific custodian of standard –Tokareva Elena  Yurvevna, s.n.s  Department 14, tei. (3952) 46-80-40,  e-mail: dep14@niiftri.irk.ru

    GET 110-2010 Standard system 1 (frequency range from 1 to 18 GHz)

    GET 110-2010 Standard systems 2, 3, 4 (frequency range from 18 to 78,33 GHz)

    Cavity resonator

    Dielectric post resonator

    Open dielectric resonator

    Open resonator

    Top